Skip to product information
1 of 9

VTSYIQI Stormer Viscometer Viscosity Meter Tester Digital rotational viscometer Rotary visometer with LCD Large Screen KU Unit 40.2 to 141.0KU Accuracy 2.00% for Paint Viscosity Testing

VTSYIQI Stormer Viscometer Viscosity Meter Tester Digital rotational viscometer Rotary visometer with LCD Large Screen KU Unit 40.2 to 141.0KU Accuracy 2.00% for Paint Viscosity Testing

Regular price $985.00 USD
Regular price Sale price $985.00 USD
Sale Sold out
Shipping calculated at checkout.

 

VTSYIQI Stormer Viscometer Viscosity Meter Tester Digital rotational viscometer Rotary visometer with LCD Large Screen KU Unit 40.2 to 141.0KU Accuracy 2.00% for Paint Viscosity Testing

Feature

Summary---The instrument is controlled by a computer. The blade of a stirred blade (also called a rotor) is driven by a constant speed motor at a speed of 200r/min. The resistance moment of the impeller in the measured sample is converted by a computer to the KU value. unit (KU) is a logarithmic function of the load required to produce 200r/min speed, which is generally used to represent the speed of the 200r/min. It is used for the viscosity of the brush and roll.
The main technical parameters---1. measurement range: 40.2 ~ 141.0KU (KU value), 27cP-5250cP 32 ~ 1099g (load value); 2. measurement accuracy: 2.0% of full range; 3. measurement repeatability: the full range of 2.0%; 4. the rotational speed of the mixing axis: 200r/min + 1r/min; 5. container capacity: About 500ml; 6. input voltage: 220V 50Hz; 7. power: 10W; 8. shape size: 210mm x 180mm x 500mm (long * width *); 9. weight: 10kg.
Operation---1. remove the instrument from the packaging box and put it on the worktable. 2. remove the stirring blade from the packing box, insert it into the sleeve of the rotating shaft, and fix it with the top tightening screw. 3. plug in the power plug, open the power switch on the back, the system starts to self check, the system will enter the standby state after the completion of the self test, and the display will display "-". 4. put the tested sample in the container
Use---6. after a few seconds, the monitor displays the viscosity value of the sample to be measured, which can be read at this time. (the sample must be stirred for a few seconds to achieve dynamic balance, and the data displayed can be considered accurate, without the display of the data before the instrument "--"). 7. after the measurement is completed, the lifting handle is lifted to the height, and there is a limited bit device near the highest position
Note---1. there are precise measuring parts in the instrument cabinet. 2. this instrument is a special instrument. Please do not open the instrument shell by itself, so as not to affect the accuracy of instrument measurement. 3. instruments are temporarily unused, the lifting handle of the instrument should be kept on the high position limit unit, and it should not be placed in a lower position, otherwise the phenomenon of inflexibility may be produced. 4. every year, companies should be check

One.Summary

The stom viscometer is used for the determination of paint and other KU indicating the viscosity of the coating. The instrument can display the KU value of the measured sample viscosity or the load value associated with the KU value (unit: G). Two.Specification

1. measurement range: 40.2 ~ 141.0KU (KU value), 27cP-5250cP 32 ~ 1099g (load value);

2. measurement accuracy: 2.0% of full range;

3. measurement repeatability: the full range of 2.0%;

4. the rotational speed of the mixing axis: 200r/min + 1r/min;

5. container capacity: About 500ml;

6. input voltage: 220V 50Hz;

7. power: 10W;

8. shape size: 210mm x 180mm x 500mm (long * width *);

9. weight: 10kg. Three.Experimental environment

1. the instrument should be placed on a strong and smooth surface away from strong airflow, strong magnetic field, strong electric field and strong electric pulse interference.

2. in order to ensure the safety of the operator, the power outlet in the laboratory should be grounded reliably.

3. the temperature in the laboratory should be 23 C + 2 C, and the relative humidity should be less than 75%.

4. sample should be 23 + 0.2 C. Four. Note

1. there are precise measuring parts in the instrument cabinet.

2. this instrument is a special instrument. Please do not open the instrument shell by itself, so as not to affect the accuracy of instrument measurement.

3. instruments are temporarily unused, the lifting handle of the instrument should be kept on the high position limit unit, and it should not be placed in a lower position, otherwise the phenomenon of inflexibility may be produced.

4. every year, companies should be checked regularly or sent to the metrological supervision department for testing once. Five.Packing list

1. host one;

2. stainless steel cup;

3.Manual

4. power line

View full details