跳至产品信息
1 / 10

VTSYIQI Digital Stormer Viscometer Viscosity Meter teter Range 40.2 to 141.0KU Accuracy 2.00% Temperature Value Ku Value g Value CP Value Time Value

VTSYIQI Digital Stormer Viscometer Viscosity Meter teter Range 40.2 to 141.0KU Accuracy 2.00% Temperature Value Ku Value g Value CP Value Time Value

常规价格 $986.00 USD
常规价格 促销价 $986.00 USD
销售额 售罄
结账时计算的运费
  • STM- IV (A) stomo viscometer specification

    ★Summary---The instrument is controlled by a computer. The blade of a stirred blade (also called a rotor) is driven by a constant speed motor at a speed of 200r/min. The resistance moment of the impeller in the measured sample is converted by a computer to the KU value. unit (KU) is a logarithmic function of the load required to produce 200r/min speed, which is generally used to represent the speed of the 200r/min. It is used for the viscosity of the brush and roll.
  • ★The main technical parameters---1. measurement range: 40.2 ~ 141.0KU (KU value), 27cP-5250cP 32 ~ 1099g (load value); 2. measurement accuracy: 2.0% of full range; 3. measurement repeatability: the full range of 2.0%; 4. the rotational speed of the mixing axis: 200r/min + 1r/min; 5. container capacity: About 500ml; 6. input voltage: 220V 50Hz; 7. power: 10W; 8. shape size: 210mm x 180mm x 500mm (long * width *); 9. weight: 10kg.
  • ★Operation---1. remove the instrument from the packaging box and put it on the worktable. 2. remove the stirring blade from the packing box, insert it into the sleeve of the rotating shaft, and fix it with the top tightening screw. 3. plug in the power plug, open the power switch on the back, the system starts to self check, the system will enter the standby state after the completion of the self test, and the display will display "-". 4. put the tested sample in the container
  • ★ Use---6. after a few seconds, the monitor displays the viscosity value of the sample to be measured, which can be read at this time. (the sample must be stirred for a few seconds to achieve dynamic balance, and the data displayed can be considered accurate, without the display of the data before the instrument "--"). 7. after the measurement is completed, the lifting handle is lifted to the height, and there is a limited bit device near the highest position
  • ★ Note---1. there are precise measuring parts in the instrument cabinet. 2. this instrument is a special instrument. Please do not open the instrument shell by itself, so as not to affect the accuracy of instrument measurement. 3. instruments are temporarily unused, the lifting handle of the instrument should be kept on the high position limit unit, and it should not be placed in a lower position, otherwise the phenomenon of inflexibility may be produced. 4. every year, companies should be check

STM- IV (A) stomo viscometer specification

One.Summary

The stom viscometer is used for the determination of paint and other KU indicating the viscosity of the coating. The instrument can display the KU value of the measured sample viscosity or the load value associated with the KU value (unit: G). Two.Specification

1. measurement range: 40.2 ~ 141.0KU (KU value), 27cP-5250cP 32 ~ 1099g (load value);

2. measurement accuracy: 2.0% of full range;

3. measurement repeatability: the full range of 2.0%;

4. the rotational speed of the mixing axis: 200r/min + 1r/min;

5. container capacity: About 500ml;

6. input voltage: 220V 50Hz;

7. power: 10W;

8. shape size: 210mm x 180mm x 500mm (long * width *);

9. weight: 10kg. Three.Experimental environment

1. the instrument should be placed on a strong and smooth surface away from strong airflow, strong magnetic field, strong electric field and strong electric pulse interference.

2. in order to ensure the safety of the operator, the power outlet in the laboratory should be grounded reliably.

3. the temperature in the laboratory should be 23 C + 2 C, and the relative humidity should be less than 75%.

4. sample should be 23 + 0.2 C. Four. Note

1. there are precise measuring parts in the instrument cabinet.

2. this instrument is a special instrument. Please do not open the instrument shell by itself, so as not to affect the accuracy of instrument measurement.

3. instruments are temporarily unused, the lifting handle of the instrument should be kept on the high position limit unit, and it should not be placed in a lower position, otherwise the phenomenon of inflexibility may be produced.

4. every year, companies should be checked regularly or sent to the metrological supervision department for testing once. Five.Packing list

1. host one;

2. stainless steel cup;

3.Manual

4. power line

Model: Digital Display Push-pull Force Meter

Maximum load value: 500N; 50kg; 110Lb; 1800Oz

Load division value:0.1N; 0.01kg; 0.01Lb; 1Oz

Sensor structure: Inner sensor

Accuracy: ±1%

查看完整详细信息